Dataset from "In silico assessment of collateral eddy current heating in biocompatible implants subjected to magnetic hyperthermia treatments"

From MaRDI portal
(Redirected from Dataset:6722521)



DOI10.5281/zenodo.4462771Zenodo4462771MaRDI QIDQ6722521FDOQ6722521

Dataset published at Zenodo repository.

Irene Rubia-Rodríguez, Luca Zilberti, Oriano Bottauscio, Mario Chiampi, Alessandro Arduino, Daniel Ortega

Publication date: 20 January 2021

Copyright license: Creative Commons Attribution 4.0 International



This dataset from the publication entitled Dataset from In silico assessment of collateral eddy current heating in biocompatible implants subjected to magnetic hyperthermia treatments contains simulated data of magnetic hyperthermia treatments for three different indications: colorectal cancer, prostate cancer and head neck cancer. Since the aim of the study is to evaluate the risk of thermal damage caused by the collateral heating of two common types of passive prostheses (hip and dental implants), eddy currents induced in these implants upon interacting with the externally applied ac field during treatment have been computed for all the evaluated regions. Two differentalloys for the implants have been considered for each case as well: Ti6Al4V and CoCrMo. At the same time, besides temperature, the specific abosorption rate (SAR) have been also computed to work out the energy deposition in tissues. Calculations have been carried out using a het exchange model with and without thermoregulation. log-log SAR vs T plots have been obtained and proposed as a quick means to pre-check treatment feasibility in each patient. These graphs are thought to be included in treatment planning prior to the clinical procedure. Other parameters taken into account have been the treatment time (5 and 30 minutes), and the maximum tolerable temperature threshold (1 or 5 C, as indicated by the ICNIRP commission), all for three main types of tissues, namely fat, bone and muscle. Each tissue have been simulated using threedifferent field intensities(5, 10 and 15 mT). The field frequency has been 300 kHz in all cases. The files Dataset_description.doc and file_scheme.txt containthe structure and description of the files that make up the dataset. UPDATES FROM PREVIOUS VERSIONS: simulations of the dental implant withoutthermoregulation have been added.







This page was built for dataset: Dataset from "In silico assessment of collateral eddy current heating in biocompatible implants subjected to magnetic hyperthermia treatments"