Summary: Geometric sensitivity for single photon emission computerized tomography (SPECT) is given by a double integral over the detection plane. It would be useful to be able to explicitly evaluate this quantity. This paper shows that the inner integral can be evaluated in the situation where there is no gamma ray penetration of the material surrounding the pinhole aperture. This is done by converting the integral to an integral in the complex plane and using Cauchy's theorem to replace it by one which can be evaluated in terms of elliptic functions.
- Theoretical determination of the collimator geometrical transfer function for the reconstruction of SPECT data with algorithms based on the numerical inversion of the Radon transform
- A spectral geometric model for Compton single scatter in PET based on the single scatter simulation approximation
- scientific article; zbMATH DE number 869709
- Analytic inversion of a conical Radon transform arising in application of Compton cameras on the cylinder
- Single-photon emission computed tomography by inverting the attenuated Radon transform with least-squares collocation