A calibration of E-field probe based on tissue-equivalent liquid waveguide and evaluation of uncertainty (Q2045579)

From MaRDI portal

!

This is the item page for this Wikibase entity, intended for internal use and editing purposes. Please use the normal view instead:

scientific article; zbMATH DE number 7381741
Language Label Description Also known as
default for all languages
No label defined
    English
    A calibration of E-field probe based on tissue-equivalent liquid waveguide and evaluation of uncertainty
    scientific article; zbMATH DE number 7381741

      Statements

      A calibration of E-field probe based on tissue-equivalent liquid waveguide and evaluation of uncertainty (English)
      0 references
      0 references
      0 references
      0 references
      0 references
      13 August 2021
      0 references
      In the present paper the authors proposed a new method to obtain the conversion factor (convF) of the E-field probe in the simulant tissue. The novelty of this proposal is that it can simplify the procedure of the conversion factor analysis, by employing the liquid waveguide with known E-filed as a component of the reference field to compare with the measured E-field of the dosimeter probe immersed in the equivalent-tissue liquid. In addition, the authors have evaluated the uncertainty of calibration for this method and establish that the result values show no excess of the standard requirement.
      0 references
      conversion factor
      0 references
      liquid waveguide
      0 references
      e-field probe
      0 references
      calibration
      0 references
      uncertainty
      0 references