Analytical model based on a cylindrical geometry to study RF ablation with needle-like internally cooled electrode
Summary: Radiofrequency (RF) ablation with internally cooled needle-like electrodes is widely used in medical techniques such as tumor ablation. The device consists of a metallic electrode with an internal liquid cooling system that cools the electrode surface. Theoretical modeling is a rapid and inexpensive way of studying different aspects of the RF ablation process by the bioheat equation, and the analytical approach provides an exact solution to the thermal problem. Our aim was to solve analytically the RF ablation transient time problem with a needle-like internally cooled cylindrical electrode while considering the blood perfusion term. The results showed that the maximal tissue temperature is reached \(\approx\) 3 mm from the electrode, which confirms previous experimental findings. We also observed that the temperature distributions were similar for three coolant temperature values (5\(^\circ\)C, 15\(^\circ\)C and 25\(^\circ\)C). The differences were only notable in temperature very close to the probe. Finally, considering the 50\(^\circ\)C line as a thermal lesion mark, we found that lesion diameter was around 2 cm, which is exactly that observed experimentally in perfused hepatic tissue and slightly smaller than that observed in nonperfused (ex vivo) hepatic tissue.
- An analytical solution for radiofrequency ablation with a cooled cylindrical electrode
- Theoretical modeling of RF ablation with internally cooled electrodes: comparative study of different thermal boundary conditions at the electrode-tissue interface
- Electrical-thermal analytical modeling of monopolar RF thermal ablation of biological tissues: determining the circumstances under which tissue temperature reaches a steady state
- A two-scale model of radio-frequency electrosurgical tissue ablation
- Modeling, simulation and optimization of radio frequency ablation
- Numerical evaluation of ablation zone under different tip temperatures during radiofrequency ablation
- Two-compartment mathematical modeling in RF tumor ablation: new insight when irreversible changes in electrical conductivity are considered
- Two-compartment mathematical modeling in RF tumor ablation: new insight when irreversible changes in electrical conductivity are considered
- An analytical solution for radiofrequency ablation with a cooled cylindrical electrode
- The radially symmetric Green's function for Dirichlet problem of pennes equation outside of a circular domain
- A fundamental solution for radiofrequency heating due to a point electrical current in an infinite spatial domain: numerical proofs with MATHEMATICA and ANSYS
- Analytical transient-time solution for temperature in non perfused tissue during radiofrequency ablation
- Effects of saline volume on lesion formation during saline-infused radiofrequency ablation
- Mathematical modeling of radiofrequency ablation for varicose veins
- Saline-enhanced RF ablation on a cholangiocarcinoma: a numerical simulation
- Mathematical 3D modelling and sensitivity analysis of multipolar radiofrequency ablation in the spine
- Theoretical modeling of RF ablation with internally cooled electrodes: comparative study of different thermal boundary conditions at the electrode-tissue interface
- Electrical-thermal analytical modeling of monopolar RF thermal ablation of biological tissues: determining the circumstances under which tissue temperature reaches a steady state
This page was built for publication: Analytical model based on a cylindrical geometry to study RF ablation with needle-like internally cooled electrode
Report a bug (only for logged in users!)Click here to report a bug for this page (MaRDI item Q1955249)